Catalog Number HAR36
Device Problems Melted (1385); Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r9398p.Investigation summary : the analysis results found that the device was received with the tissue pad detached and not returned and with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.Due to the returned condition of the device, we were unable to further investigate the tissue effect issues reported.The device was disassembled to inspect the internal components and no anomalies were found.Based on the condition of the tissue pad, a probable cause for this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.The batch history record was reviewed, and no defects, nc¿s, or protocols related to the complaint were found during the manufacturing process.
 
Event Description
It was reported that during a hiatal herniorrhaphy procedure, the device did not seal and cut normally, it took a long time.Finally the white plastic melted by having the caliper so long in use.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC LAP 5MM SHEAR 36CM
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7830301
MDR Text Key118749478
Report Number3005075853-2018-12344
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014444
UDI-Public20705036014444
Combination Product (y/n)N
PMA/PMN Number
K120729
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 08/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Catalogue NumberHAR36
Device Lot NumberR9398P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/17/2018
Initial Date FDA Received08/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/18/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-