The detachment of the covering was confirmed.Four adhesive spots are visible on the covering.The od of the stent when shipped is 0.181".The od of this stent is 0.114", and there is significant twisting in the zig rows at each end of the stent.As stated in the report from the hospital, they were attempting to use this device in a 9 fr.Introducer.In the instructions for use provided with the device, it states that this device when mounted on a 12mm bib required a 12 fr.Introducer.It also states in the report that they crimped this device on the bib catheter with umbilical tape.This is not the recommended method that is stated in the instructions for use.Between crimping this device with umbilical tape and trying to pass it into an introducer 3 sizes smaller than the rated introducer size, most likely caused the covering to detach.A sample from each lot of stents is tested for covering adhesion strength.The sample from this lot failed at 3.01lbs, which is well above the 1.5lb minimum acceptance criteria.
|