MAUDE Adverse Event Report: ABBOTT DIAGNOSTICS DIVISION CELL-DYN SAPPHIRE ANALYZER; AUTOMATED HEMATOLOGY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIAGNOSTICS DIVISION CELL-DYN SAPPHIRE ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 08H00-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/11/2018
Event Type  malfunction  
Manufacturer Narrative
All available patient information was included.Additional patient details are not available.An evaluation is in process.A final report will be submitted when the evaluation is complete.
 
Event Description
The customer observed a falsely elevated platelet result for one patient on the cell-dyn sapphire analyzer.The following data was provided for a diagnosed with bile obstruction and taking multiple medications: (b)(6).There was no impact to patient management reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs and a review of labeling.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The instruments generated lower, upper or lower and upper region interference in plti histogram on all the data provided and some runs has additional flag, pic/poc delta.All the runs has invalid data, asterisk on mpv results and some are on plto and plti.Review of numerical data, data flags and scatterplots is consistent with the presence of an interfering substance.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN SAPPHIRE ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key7841968
MDR Text Key119666051
Report Number2919069-2018-00044
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740016616
UDI-Public00380740016616
Combination Product (y/n)N
PMA/PMN Number
K051215
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H00-01
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 08/14/2018
Initial Date FDA Received09/04/2018
Supplement Dates Manufacturer Received09/20/2018
Supplement Dates FDA Received09/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-