• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500 Back to Search Results
Model Number 500AHCT
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 08/13/2018
Event Type  malfunction  
Manufacturer Narrative
The srt replaced the arterial bpm.The unit operates to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The service repair technician (srt) reported that during routine testing of the device at the service center, the arterial blood parameter monitor (bpm) had failed the high potential (hi-pot) test.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 500
Type of Device
MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7845975
MDR Text Key119497213
Report Number1828100-2018-00459
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
PMA/PMN Number
K133658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number500AHCT
Device Catalogue Number500AHCT
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/13/2018
Initial Date FDA Received09/05/2018
Supplement Dates Manufacturer Received09/25/2018
Supplement Dates FDA Received09/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-