On (b)(6) 2018, patient (b)(6) at the university (b)(6) was unable to charge the device and had a flashing red light on the mc.The reshape lifesciences field specialist connected her to the cp and alarm codes #4 and #25 were observed.These alarms were cleared and the flashing red light disappeared.The cp continued to ask to start a charge as the rnr battery was extremely low.After about 15 minutes of charging using the patients mc, therapy was able to be adjusted and a live test performed.On-demand impedance measurement was performed and the results were as follows: 934, 1033, 677.Before the patient left, another on-demand impedance test was performed, the readings were as follows: 821, 1062, 0.A red flashing light was observed on the mc.The patient was hooked up to the cp to clear the red light and was sent home.On (b)(6) 2018, the patient indicated they received another flashing red light.
|