• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITRON SELECT SPS ULTRASONIC SCALER; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITRON SELECT SPS ULTRASONIC SCALER; SCALER, ULTRASONIC Back to Search Results
Model Number G124
Device Problems Restricted Flow rate (1248); Overheating of Device (1437); Blocked Connection (2888)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While using a g124 scaler, the handpiece was clogged and overheating; no injury resulted.
 
Manufacturer Narrative
Unit is beyond useful life.The handpiece is clogged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITRON SELECT SPS ULTRASONIC SCALER
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
MDR Report Key7852663
MDR Text Key119651931
Report Number2424472-2018-00134
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
PMA/PMN Number
K970123
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG124
Device Catalogue Number81324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Initial Date Manufacturer Received 08/07/2018
Initial Date FDA Received09/06/2018
Supplement Dates Manufacturer Received10/17/2018
Supplement Dates FDA Received11/14/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-