MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Information (3190)
Event Date 03/22/2018
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2 mm vticmo13.2, -8.5/1.5/85 (sphere/cylinder/axis), implantable collamer lens into the patient's right eye (od) on (b)(6) 2018.On (b)(6) 2018 the lens was exchanged for a shorter lens due to observation of excessive vault.This exchange resolved the problem.
 
Manufacturer Narrative
Device evaluation: lens was returned dry in a micro centrifuge vial.There was clear surgical residue on product.Visual inspection found no visible damage to the lens and residue on the lens surface.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7855444
MDR Text Key119529856
Report Number2023826-2018-01353
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model NumberVTICMO13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received09/07/2018
Supplement Dates Manufacturer Received10/09/2018
Supplement Dates FDA Received10/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age22 YR
-
-