STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.2
Device Problems Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923)
Patient Problems No Information (3190); No Code Available (3191)
Event Date 03/02/2018
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 13.2 mm vticmo13.2, -7.5/4/75 (sphere/cylinder/axis), implantable collamer lens into the patient's left eye (os) on (b)(6) 2017.On (b)(6) 2018 the lens was exchanged for a longer lens due to observation of low vault.This exchange resolved the problem.
 
Manufacturer Narrative
Additional data: device evaluation: lens was returned dry in a micro centrifuge vial.There was clear surgical residue on product.Visual inspection found the haptic torn and residue on the lens surface.Corrected data: original mdr stated "on 06/07/18 the lens was exchanged" which should be corrected to "on (b)(6) 2018 the lens was explanted due to observation of low vault with lens rotation.On (b)(6) 2018 a longer length lens was implanted.The problem was resolved." (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7856947
MDR Text Key119575699
Report Number2023826-2018-01356
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model NumberVTICMO13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received09/07/2018
Supplement Dates Manufacturer Received10/10/2018
Supplement Dates FDA Received10/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age29 YR
-
-