Model Number VICMO12.6
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problems No Information (3190); No Code Available (3191)
Event Date 07/04/2018
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 12.6mm vicmo12.6, -9.00 diopter, implantable collamer lens into the patient's right eye (od) on (b)(6) 2018.On (b)(6) 2018 the lens was exchanged for a longer lens due to observation of low vault.This exchange resolved the problem.
 
Manufacturer Narrative
Device evaluation: lens was returned dry in a micro-centrifuge vial.There was clear surgical residue on product.Visual inspection found a deformed haptic and presence of residue on lens surface.(b)(4).
 
Manufacturer Narrative
Corrected data: device evaluation: visual inspection- "lens was caught on lens vial cap" not included in supplement mdr#1.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7857161
MDR Text Key119589479
Report Number2023826-2018-01357
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Model NumberVICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received09/07/2018
Supplement Dates Manufacturer Received10/10/2018
11/01/2018
Supplement Dates FDA Received10/31/2018
11/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age43 YR
-
-