• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICMO12.6
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problems No Information (3190); No Code Available (3191)
Event Date 04/03/2018
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that the surgeon implanted a 12.6mm vicmo12.6, -8.50 diopter, implantable collamer lens into the patient's right eye (od) on (b)(6) 2017.On (b)(6) 2018 the lens was exchanged for a longer lens due to observation of low vault.This exchange resolved the problem.
 
Manufacturer Narrative
Additional data: device evaluation: lens was returned dry in a micro-centrifuge vial with clear surgical residue on product.Visual inspection found the haptic torn with foreign material (residue) on lens surface.Corrected data: the reporter indicated that the surgeon implanted a 12.6mm vicmo_12.6 implantable collamer lens (-8.5 diopter) into the patient's right eye (od) on (b)(6) 2017.The lens was explanted due to low vault on (b)(6) 2018.A longer replacement lens was implanted in a later surgery on (b)(6) 2018; problem was resolved.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7857365
MDR Text Key119589234
Report Number2023826-2018-01358
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model NumberVICMO12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/08/2018
Initial Date FDA Received09/07/2018
Supplement Dates Manufacturer Received10/10/2018
Supplement Dates FDA Received10/28/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age27 YR
-
-