Product evaluation: failure analysis for fiber 0010-2090-21735927-(b)(4): the glass cap exhibits devitrification, and slight detritus buildup around the devitrification area; the glass cap is fractured at the fiber/glass cap fusion zone; the distal portion of glass cap fracture is missing; the heat shrink tubing open end exhibits minor scratch marks.Based on the device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: operational context.The probable root cause was determined based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.
|