• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT HPS BPH FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT HPS BPH FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2090
Device Problems Fracture (1260); Detachment of Device or Device Component (2907); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/10/2018
Event Type  malfunction  
Manufacturer Narrative
Product evaluation: failure analysis for fiber 0010-2090-21735927-(b)(4): the glass cap exhibits devitrification, and slight detritus buildup around the devitrification area; the glass cap is fractured at the fiber/glass cap fusion zone; the distal portion of glass cap fracture is missing; the heat shrink tubing open end exhibits minor scratch marks.Based on the device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: operational context.The probable root cause was determined based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.
 
Event Description
It was reported that during a benign prostatic hyperplasia surgical procedure at 176,902 joules and 31:38 minutes fiber tip damage was noted.The fiber was exchanged and the case was completed using an alternate procedure ("electric knife").The fiber tip was cleaned during the procedure.Good patient outcome reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT HPS BPH FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7858547
MDR Text Key119646551
Report Number2937094-2018-00983
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953004709
UDI-Public00878953004709
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/22/2020
Device Model Number0010-2090
Device Catalogue Number0010-2090
Device Lot Number21735927
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/30/2018
Initial Date FDA Received09/08/2018
Supplement Dates Manufacturer Received09/19/2018
Supplement Dates FDA Received10/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-