Product evaluation: the glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone of the bevel edge; the fiber proximal to fracture can rotate independently of outer flow tubing; the glass cap exhibits severe devitrification at output window; the metal cap exhibits mild detritus adhesion; the outer flow tubing open end exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is: operational context based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.
|
It was reported that during a benign prostatic hyperplasia surgical procedure at 15,961 joules and 1:40 minutes decreased vaporization effectiveness was noted.The fiber was exchanged and at 9,740 joules and 1:45 minutes the same problem was noted.The case was completed using an alternative procedure (bipolar resection).The tip of the first fiber was cleaned during the procedure.No harm to the patient reported.This report is for the second fiber.
|