Model Number 0010-2400
Device Problems Fracture (1260); Temperature Problem (3022); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/17/2018
Event Type  malfunction  
Event Description
It was reported that during a prostate procedure, forward firing occurred after 56,243j with 10:17 minutes of fiber use.The tip of the fiber was cleaned during procedure.The case was successfully completed with a second fiber without clinical consequences to the patient.
 
Manufacturer Narrative
Analysis of the device revealed the glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge.The fiber proximal to fracture can rotate independently of outer flow tubing.The glass cap exhibits severe devitrification at output window.The metal cap exhibits mild char on surface.The outer flow tubing open end exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.The probable cause was determined based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.An evaluation conclusion code of cause cannot be traced to device was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7862561
MDR Text Key119791275
Report Number2937094-2018-00994
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/23/2019
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number726B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/27/2018
Initial Date FDA Received09/10/2018
Supplement Dates Manufacturer Received09/14/2018
Supplement Dates FDA Received10/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-