AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2400
Device Problems Fracture (1260); Detachment of Device or Device Component (2907); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/20/2017
Event Type  malfunction  
Event Description
It was reported that during a benign prostatic hyperplasia (bph) procedure, with 153,353 joules and 14:56 minutes used, it was noted that the metal tip detached from the fiber.The procedure was complete utilizing a second fiber (another of the same device).The patient's condition following procedure was reported as stable.
 
Manufacturer Narrative
Product evaluation: failure analysis for fiber (b)(4): the fiber tip is attached to fiber; the glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge; the fiber proximal to fracture can rotate independently of outer flow tubing; the glass cap exhibits severe devitrification at output window; the metal cap exhibits severe char on surface; the outer flow tubing open end exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.The reported issue of tip detachment could not be confirmed.Probable cause: the probable root cause of distal circumferential fracture is operational context based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7862783
MDR Text Key119791271
Report Number2937094-2018-00992
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953005515
UDI-Public00878953005515
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/23/2018
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number648A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/05/2018
Initial Date FDA Received09/10/2018
Supplement Dates Manufacturer Received09/26/2018
Supplement Dates FDA Received09/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-