Catalog Number 720054-02
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 08/17/2018
Event Type  Injury  
Manufacturer Narrative
The analysis results indicate one cylinder performed within specifications.The other cylinder had a hole in the outer layer that was consistent with damage caused by a sharp instrument that exposed the inner segments.This cylinder functioned as intended.
 
Event Description
It was reported that the patient had his spectra concealable penile prosthesis replaced with a 3 piece inflatable penile prosthesis due to dissatisfaction."patient complained the rigidity of prosthesis is not good".No patient complications were reported in relation with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS SPECTRA CONCEALABLE PENILE PROSTHESIS
Type of Device
PROSTHESIS, PENILE
Manufacturer (Section D)
BOSTON SCIENTIFIC - MINNETONKA
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
BOSTON SCIENTIFIC - MINNETONKA
10700 bren road w
minnetonka MN 55343
Manufacturer Contact
sharon zurn
10700 bren road w
minnetonka, MN 55343
9529306000
MDR Report Key7863615
MDR Text Key119783466
Report Number2183959-2018-00110
Device Sequence Number1
Product Code FAE
UDI-Device Identifier00878953005218
UDI-Public00878953005218
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 08/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/17/2015
Device Catalogue Number720054-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/20/2018
Initial Date FDA Received09/11/2018
Date Device Manufactured12/01/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age65 YR
-
-