• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE; BARD URETERAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE; BARD URETERAL STENT Back to Search Results
Model Number 777624
Device Problems Break (1069); Material Split, Cut or Torn (4008)
Patient Problems No Patient Involvement (2645); No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the end of the guidewire had a split.
 
Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "bard® guidewires information for use description: guidewires are used to provide access and facilitate passage of endourological instrumentation during urological procedures.They are offered in many different configurations that include: sizes (lengths and diameters), tip designs, coatings (hydrophilic and ptfe) and stiffness.The bard® nicore¿ wire has a nitinol core and a hydrophilic coating.The bard® hydro-glide¿ stainless steel guidewires are available with a hydrophilic coating.The bard® stainless steel guidewires are available with and without ptfe and are available in fixed and movable core.Indications: bard® guidewires are indicated to provide transurethral and/or percutaneous access into the bladder, ureter or renal pelvis.Contraindications: there are no known contraindications warning: inspect all guidewires for damage prior to use.Bending or kinking during or prior to placement could damage the guidewire.Do not attempt to use the guidewire if it has been damaged.Use of a damaged wire may result in damage to the urinary tract.Do not reshape the guidewire by any means.Attempting to reshape the guidewire may cause damage resulting in release of fragments into the urinary tract.Failure to exercise proper caution may result in damage to the urinary tract.Do not manipulate or remove the guidewire through a metal cannula or needle.This may result in destruction/separation of the outer jacket of the wire requiring retrieval.Use extreme caution when using a laser, making sure to avoid contact with the wire.Direct contact could result in damage / breakage to the wire.Attention should be paid to guidewire movement in the urinary tract.Before a guidewire is moved or torqued, tip movement should be examined under direct vision or fluoroscopy.Do not advance or withdraw a guidewire when resistance is encountered as perforation could occur.Sufficient guidewire length must remain exposed to maintain a firm grip on guidewire at all times.Failure to comply with the warnings could result in damage to the guidewire to include, but not limited to: wire breakage, abrasion of the coating, release of guidewire fragments into the urinary system, all of which might require intervention.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Precautions: should be used only by physicians thoroughly trained in endourological guidewire techniques.Do not use dry gauze to manipulate the guidewire as this can damage the surface coating and make the wire tacky.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Potential complications: complications that may result from the use of a guidewire in a urological procedure incude, but are not limited to: ¿ perforation ¿ acute or delayed bleeding ¿ tissue trauma ¿ edema direction for use: the physician should select the proper size and length of the guidewire for the procedure being performed.1.The guidewire is packaged in a protective coil.For hydrophilic coating wires: hydrophilic coated guidewires require activation of their coating.Prior to removing the guidewire from the protective coil, inject sterile saline through the port to activate the lubricious coating.Remove the guidewire from the protective coil and carefully inspect the coil for separation, bends, kinks or a damaged tip.For ptfe wires: remove the guidewire from the protective coil and carefully inspect the coil for separation, bends, kinks or damaged tip.2.Introduce the guidewire, flexible end first, into the working channel of the endoscopes or percutaneously into the urinary tract.3.Advance the guidewire slowly into the desired position.Continuously confirm guidewire position either visually or under fluoroscopy.4.Carefully withdraw the guidewire taking care not to kink.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and with applicable laws and regulations.References: 1.Clayman, r.V., castaneda-zuniga, w., techniques in endourology: a guide to the percutaneous removal of renal and ureteral calculi, copyright 1984 by ralph v.Clayman.2.Smith, a.D., the urologic clinics of north america: symposium on endourology, w.B.Saunders co., february, 1982.3.Ring, e.J., mclean, g.K., interventional radiology: principles and techniques, little, brown and company, october, 1981.".
 
Event Description
It was reported that the end of the guidewire had a split.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Type of Device
BARD URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7863894
MDR Text Key119933415
Report Number1018233-2018-04096
Device Sequence Number1
Product Code FAD
UDI-Device Identifier10801741014540
UDI-Public(01)10801741014540
Combination Product (y/n)N
PMA/PMN Number
K983498
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup
Report Date 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model Number777624
Device Catalogue Number777624
Device Lot NumberNGBV2185
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/21/2018
Initial Date FDA Received09/11/2018
Supplement Dates Manufacturer Received09/26/2018
Supplement Dates FDA Received10/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-