Model Number N/A
Device Problem Malposition of Device (2616)
Patient Problem Joint Dislocation (2374)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr's were filed for this event.Please see associated report(s): 0001825034-2018-04303.0001825034 - 2018 - 04302.Reported event was confirmed via x-ray report.Review of provided patient x-rays confirms the reported implant dislocation as well as varus positioning of the femoral stem.Review of device history records was not possible as the necessary product/lot code combination(s) was not provided.Root cause was unable to be determined due to limited information received from the customer.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Not returned to manufacturer.
 
Event Description
It was reported the patient received multiple closed reductions to address chronic lateral femoral head dislocation from the acetabular construct.The implants remain in-vivo and no revision has been scheduled to date.No further information is available at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN STEM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7869221
MDR Text Key119975172
Report Number0001825034-2018-08817
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/07/2018
Initial Date FDA Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNKNOWN PART/LOT, LINER, HEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age82 YR
-
-