• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHATHERM NEPTUNE, DUAL THERMISTOR, A; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHATHERM NEPTUNE, DUAL THERMISTOR, A; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Catalog Number 395-90
Device Problem Device Sensing Problem (2917)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2018
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4.A visual and functional inspection of the device involved in the complaint could not be conducted since the device was not returned at the time of this report.However, (b)(4)devices were taken from the current production p/n 395-90 conchatherm neptune, dual thermistor, ad lot # 73h1800167.The samples were visually inspected.During testing, the issue reported was not observed in the current manufacturing process.A device history record investigation did not show issues related to this complaint.A record assessment (fmea) was performed and no update is required.Customer complaint cannot be confirmed due the device sample is not available to perform a proper investigation and determine the root cause.Corrective actions cannot be established.If the device sample becomes available at a later date, this report will be updated accordingly.
 
Event Description
Customer complaint alleges the device probe was not functioning during a patient use.No patient harm reported.Patient condition reported as fine.
 
Manufacturer Narrative
(b)(4).One (1) 395-90 conchatherm neptune, dual thermistor, ad was received for investigation.A visual examination was performed to determine if the probe sustained any abuse/misuse/damage.No damage was noted.The temperature probe was connected to a known functioning neptune heater 425-00 (serial # (b)(4)).The temperature sensors were inserted into the temperature ports of an 880-36kit breathing circuit.The temperature probe set was connected to the neptune heater via the three pin plug connection point on the lower front face of the neptune heater.The neptune heater was turned on and immediately displayed a failed temperature reading along with an audible alarm.The suspect probe was removed and a known functioning temperature probe was connected to the neptune heater.Upon restarting the neptune, the heater moved through all of the self-tests and diagnostics without interruption and moved to the operation setup mode.Based on the investigation performed, the reported complaint has been confirmed.The 395-90 temperature probe is inoperable.While the temperature probe has been confirmed as inoperable, the root cause cannot be established at this time.This product is a purchased finished goods product and will be sent to the manufacturer for further evaluation.
 
Event Description
Customer complaint alleges the device probe was not functioning during a patient use.No patient harm reported.Patient condition reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHATHERM NEPTUNE, DUAL THERMISTOR, A
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
MDR Report Key7873216
MDR Text Key120245825
Report Number3003898360-2018-00750
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K063758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number395-90
Device Lot NumberSN:616121083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Initial Date Manufacturer Received 08/30/2018
Initial Date FDA Received09/13/2018
Supplement Dates Manufacturer Received10/10/2018
Supplement Dates FDA Received10/12/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NONE REPORTED.; NONE REPORTED.
-
-