MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. PITUITARY RONGEUR BLACK 3MM LATERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. PITUITARY RONGEUR BLACK 3MM LATERAL Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/15/2018
Event Type  malfunction  
Manufacturer Narrative
Udi number: na.Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that the tip of a rongeur broke off during surgery.An alternative instrument was used to complete the procedure.There was no adverse patient events or delay of surgery.
 
Manufacturer Narrative
Additional information: method, results, and conclusions - the returned device was examined.The jaw was confirmed to have fractured off.The cause cannot be determined, but similar events have found that the jaws have fractured when high forces were applied, as when the jaw was used to cut bone instead of the disc material.There were no manufacturing issues detected which would have contributed to this event.The labeling was reviewed and found to contain instructions regarding device usage.
 
Event Description
It was reported that the tip of a rongeur broke off during surgery.An alternative instrument was used to complete the procedure.There was no adverse patient events or delay of surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PITUITARY RONGEUR BLACK 3MM LATERAL
Type of Device
RONGEUR
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
ashley mcpherson
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7875206
MDR Text Key120242795
Report Number3012447612-2018-00762
Device Sequence Number1
Product Code HTX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number8733-7153
Device Lot NumberIM00242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/16/2018
Initial Date FDA Received09/13/2018
Supplement Dates Manufacturer Received01/09/2019
Supplement Dates FDA Received01/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/21/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-