The bwi failure analysis lab received the device for evaluation.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.(b)(4).
|
It was reported that a patient underwent an ablation procedure with a thermocool® smart touch¿ electrophysiology catheter and a deflection issue occurred.The investigational analysis has been completed.The device was inspected and white material was found stuck in the tip electrode and ring #1.Then, deflection testing was performed and the catheter failed.A failure analysis was performed and the catheter handle was opened.The puller wire was found broken, a closer inspection revealed that the polymide was bonded to the side arm assembly causing the deflection issue.Fourier transform infrared spectroscopy testing (ftir) was performed and the results showed that the material was not associated with the human-like material.No similar composition was found in the device.Based on bands, foreign material is primary composed of an organic salt; presumably a carbonate.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.During manufacturing process, all the catheters are inspected for visual damages before packaging.On line inspections and functional tests are in place to prevent this type of failure from leaving the facility.The customer complaint was confirmed.The root cause of the foreign material observed cannot be determined since there is evidence that the device was manufactured in accordance with documented specification and procedures.Manufacture ref.No: (b)(4).
|