Results: the 3maxc was ovalized approximately 100.0 cm from the hub.Unraveled metal wire was protruding from the hub.Bends were present along the length of the device.Conclusions: evaluation of the returned psr3d revealed an undamaged device.No visible damage was observed on the psr3d and it was able to be advanced through a demonstration catheter without issue.The reported complaint could not be confirmed.Evaluation of the returned 3maxc revealed coil winds unraveled and protruding from its hub.If the catheter liner becomes compromised, the psr3d may interact with the coil winds upon retraction.The coil winds were stuck on the distal end of the psr3d.The subsequent retraction of the psr3d likely resulted in the unraveling of the 3maxc inner coil winds.Bends were present along the length of the 3maxc and the psr3d delivery wire.Based on the returned condition and the reported complaint, the bends are likely incident reported complaint and may have occurred during packaging for return to penumbra.Penumbra catheters and revascularization devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure in the middle cerebral artery (mca) using a penumbra system 3d revascularization device (psr3d) and a penumbra system 3max reperfusion catheter (3maxc).During the procedure, the psr3d broke apart and stretched when advanced only slightly into the 3maxc during the first pass and, therefore, the 3maxc was removed with the psr3d inside.The procedure was completed using a new psr3d and a new 3maxc.There was no report of an adverse effect to the patient.
|