• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE PLUS 7 W ADV HEMOSTASIS; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARH45
Device Problem Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p94f61.Maude event # mw5077346.Investigation summary: the device was received with the tissue pad detached and not returned and with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.The instrument was disassembled to inspect the internal components and no evidence was found that could have caused the reported activation issues.Based on the condition of the tissue pad, a probable cause of this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
 
Event Description
It was reported that during an unknown procedure, the harmonic device failed to cut and coagulate while dissecting through tissue.A new instrument worked correctly and procedure was completed without further incident.There were no patient consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 W ADV HEMOSTASIS
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7879136
MDR Text Key120529109
Report Number3005075853-2018-12769
Device Sequence Number1
Product Code LFL
UDI-Device Identifier20705036014635
UDI-Public20705036014635
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 08/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Catalogue NumberHARH45
Device Lot NumberP92M92
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 08/31/2018
Initial Date FDA Received09/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2017
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-