AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2400
Device Problems Fracture (1260); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2018
Event Type  malfunction  
Manufacturer Narrative
Product evaluation: failure analysis for fiber (b)(4): the glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge; the fiber proximal to fracture can rotate independently of outer flow tubing; the glass cap exhibits severe devitrification at output window; the metal cap exhibits mild char on surface; the outer flow tubing open end exhibits minor scratch marks.Based on device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the probable root cause of the failure is operational context based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.
 
Event Description
It was reported during bph procedure three fibers failed.The first fiber failed due to fiber timing or dying out at 200,000 joules and 35 mins of use.The fiber tip (cap) was not cleaned during the procedure.The second fiber failed due to fiber going into standby after 10,000 joules of use.The third fiber also exhibited problems but the procedure was completed using third fiber.Patient outcome: "ok" reported.No injury to patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7880931
MDR Text Key120527017
Report Number2937094-2018-01027
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953005515
UDI-Public00878953005515
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other
Type of Report Initial
Report Date 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/21/2020
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number21885134
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2018
Initial Date Manufacturer Received 09/13/2018
Initial Date FDA Received09/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-