• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM; EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION COMPACT
Device Problems Display or Visual Feedback Problem (1184); Intermittent Communication Failure (4038)
Patient Problem No Patient Involvement (2645)
Event Date 08/29/2018
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could be replicated.To restore functionality, the navigation system was replaced.The navigation system was returned to the manufacturer for analysis.The system was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Event Description
Medtronic received information that, while in a functional endoscopic sinus surgery (fess), the display on the monitor of the navigation system was described to be "fuzzy" with artifacts at start up.Restarting restored functionality temporarily as the issue would recur at the registration portion of the application software.Restarting eventually led to the navigation system not restarting.Following the procedure, when the issue was reported to medtronic, the navigation system was functioning as designed.There was no reported delay to the procedure due to this issue.There was no reported impact on patient outcome.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC FUSION COMPACT NAVIGATION SYSTEM
Type of Device
EAR NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7886865
MDR Text Key120694804
Report Number1723170-2018-04796
Device Sequence Number1
Product Code PGW
UDI-Device Identifier00643169672956
UDI-Public00643169672956
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153247
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFUSION COMPACT
Device Catalogue Number9735602
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/29/2018
Initial Date FDA Received09/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age17 YR
Patient Weight71
-
-