• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. VIDEOARTHROSCOPE HD 4MM X 30 DEG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. VIDEOARTHROSCOPE HD 4MM X 30 DEG Back to Search Results
Catalog Number 72202961S
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2018
Event Type  malfunction  
Event Description
It was reported that the device has spots on lens.There was no backup available.No significant delay or patient injury reported.
 
Manufacturer Narrative
An evaluation was performed by the supplier and could confirm the customer complaint for the device has spots on the lens.A visual inspection was performed and showed the scope to have distal tip damage and a dented outertube.This damage is caused by contact with another source.No manufacturing related defects were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOARTHROSCOPE HD 4MM X 30 DEG
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
MDR Report Key7891320
MDR Text Key120973514
Report Number3003604053-2018-00150
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
PMA/PMN Number
K043395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72202961S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2018
Initial Date Manufacturer Received 08/31/2018
Initial Date FDA Received09/19/2018
Supplement Dates Manufacturer Received10/26/2018
Supplement Dates FDA Received10/29/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-