• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP-JAPAN
Device Problems Inability to Irrigate (1337); Unstable (1667)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A physician reported that the patients intraocular pressure became unstable and irrigation failure occurred during the procedure.The product was replaced and the procedure was completed.There was no patient harm.The sample is not available for evaluation as it was discarded by the facility.
 
Manufacturer Narrative
The lot complaint history was reviewed, this is the third complaint for the finished goods lot and second for this issue for this lot.The device history record (dhr) shows the product was released per specifications.No sample was received.Therefore, no visual inspection or functional testing could be conducted.The root cause of the customer's complaint could not be established as a sample has not been received.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.After an investigation of this complaint, it has been determined that no action will be taken at this time as a sample was not returned and no root cause could be established.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Consumables manufacturing management has been made aware of this complaint through the consumer affairs review meeting.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key7899620
MDR Text Key121566381
Report Number2028159-2018-02011
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/06/2018
Initial Date FDA Received09/21/2018
Supplement Dates Manufacturer Received12/05/2018
Supplement Dates FDA Received12/07/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION SURGICAL PROCEDURE PAK
-
-