Model Number 1174108D
Device Problems Difficult to Insert (1316); Material Integrity Problem (2978); Material Twisted/Bent (2981)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/22/2018
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: two 3cg stylets were returned for investigation.It was implicated that the first stylet was used on a patient and the second stylet was used as a reference to measure the first stylet.Both stylets had been removed from the t-lock extension set.The product ifu states, ¿withdraw the entire t-lock connector/stylet assembly, as one unit.¿ a complete break was observed in the magnetic section of each stylet.A segment of polyimide tubing was missing from each stylet.On the first stylet, the polyimide tubing was received in two segments.The proximal segment was affixed to the core wire and was 2.7cm in length.The polyimide tubing was kinked 1.4cm from the proximal end of the tubing that was adhered to the core wire.The distal segment of the polyimide tubing was 2.0 cm in length and contained the conductive epoxy at the distal tip.A 2cm segment of core wire extended from the fractured end of the polyimide tubing at the proximal end of the distal stylet segment.A section of the polyimide tubing and magnets were missing from the middle of the stylet¿s magnetic tip.The core wire was kinked 2cm and 30cm from the distal end of the yellow tab.On the second stylet, the polyimide tubing was 5.2cm in length.A break was observed in the polyimide tubing and the distal segment of the stylet was not returned for investigation.Kinks were observed in the polyimide tubing.A microscopic examination revealed that the broken ends of the polyimide tubing were noted to be rough and uneven.Blood residue was observed within the polyimide tubing of the used sample.The wall thickness of the polyimide tubing was within specification for each sample.Due to the observed kinks and fractures in the polyimide tubing, the complaint was confirmed and the cause was determined to be related to use of the device.Placement of the catheter where kinking may occur should be avoided.Kinking of the stylet may result in stylet damage.A lot history review (lhr) of rect0834 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the rn inserted picc, unscrewed top and pulled wire back and it was stated it felt tight.Then the rn pulled the picc a little ways, and then reinserted the wire and lure locked the wire back in.The rn then inserted picc to svc, obtained cxr and inspected the wire post insertion.It was stated the end of the 3cg wire was bent in two places post insertion.The rn notice two weakened spots near the end of the wire (distal).It was said the wire sheared as the rn pulled it through ¿tougey¿.Returned were two wires, one was used to measure the reported wire.This report addresses the reported sheared wire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET 4F
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
kelsey erickson
605 n. 5600 w.
salt lake city, UT 84116
8015225937
MDR Report Key7899756
MDR Text Key121580699
Report Number3006260740-2018-02548
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741034459
UDI-Public(01)00801741034459
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091324
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1174108D
Device Catalogue Number1174108D
Device Lot NumberRECT0834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 09/04/2018
Initial Date FDA Received09/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-