Model Number P005723CAP02
Device Problem Inflation Problem (1310)
Patient Problem Pressure Sores (2326)
Event Date 08/31/2018
Event Type  Injury  
Manufacturer Narrative
The account stated that the mattress was not inflating properly and the patient had a pre existing wound on his right ischium that was almost healed, but has now progressed into a stage 3 wound.The patient stated that he is receiving the same daily wound dressing changes, but more frequently due to the extra leakage.A stage 3 pressure ulcer meets the criteria for a serious injury per fda definition.Development of pressure ulcers is multi-factorial and cannot be only attributed to performance of the surface.Risk factors include protein-calorie malnutrition, micro-climate (skin wetness caused by sweating or incontinence), diseases that reduce blood flow to the skin, such as arteriosclerosis, or diseases that reduce the sensation in the skin, such as paralysis or neuropathy.Position changes are key to pressure sore prevention and treatment.These changes need to be frequent, re-positioning needs to avoid stress on the skin, and body positions need to minimize the risk of pressure on vulnerable areas.The hill-rom technician evaluated the mattress and blower and found the hose connections were clogged due to nicotine build up.He also found some of the hose connections were broken and unable to connect due to the damaged sustained from the account moving the unit from room to room.The blower unit was also showing a pneumatic error 3 due to the air loss.It is necessary for the p500 to have an effective maintenance program.We recommend that you do annual preventive maintenance.Examine the air filter for cleanliness.Clean or replace as necessary.Make sure the hose, hose connectors, hose receptacles, and end covers are not damaged (no kinks, cracks, holes, broken parts, etc.) replace or repair as necessary.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the patient stating his pre existing wound was almost healed and now it has gotten worse.The blower and mattress are located at the patients home.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P500 MRS CONTROL UNIT
Type of Device
BED, FLOTATION THERAPY, POWERED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key7905462
MDR Text Key121583153
Report Number1824206-2018-00372
Device Sequence Number1
Product Code IOQ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberP005723CAP02
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2018
Initial Date FDA Received09/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/24/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age34 YR
Patient Weight91
-
-