• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN HEAD; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN HEAD; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Material Erosion (1214)
Patient Problem Local Reaction (2035)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Journal article: waterson, h.B., whitehouse, m.R., greidanus, n.V., garbuz, d.S., masri, b.Z., duncan, c.P.(2018).Revision for adverse local tissue reaction following metal-on-polyethylene total hip arthroplasty is associated with a high risk of early major complications.The british editorial society of bone & joint surgery.100-b:720¿4.Https://doi.Org/10.1302/0301-620x.100b6.Bjj- 2017-1466.R1.Concomitant medical products: item# unknown, unknown stem, lot# unknown; item# unknown, unknown liner, lot# unknown; item# unknown, unknown cup, lot# unknown.Foreign source: (b)(6).Multiple mdr reports were filed for this event; please see associated reports: 0001822565-2018-05209.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported in journal article that nine patients presented with elevated metal ions preoperatively.Patient proceeded with revision of tha.During the procedure the joint was found to have developed armd, confirmed by intraoperative pathology samples which revealed necrotic granulomatous and trunnionosis that was identified by visible thick black debris on the trunnion.Attempts were made to obtain additional information; however, none was available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HEAD
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7905557
MDR Text Key121585501
Report Number0001822565-2018-05210
Device Sequence Number1
Product Code JDG
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
PNI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/30/2018
Initial Date FDA Received09/25/2018
Supplement Dates Manufacturer Received10/18/2018
Supplement Dates FDA Received11/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-