Model Number VTICMO13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Code Available (3191)
Event Date 06/21/2018
Event Type  Injury  
Manufacturer Narrative
This product is not marketed in the us.(b)(4).
 
Event Description
The reporter indicated that a 13.2mm, vticmo13.2, -14.5/+2.0/ 83 (sphere/cylinder/axis), implantable collamer lens was implanted into the patients right eye (od) on (b)(6) 2018.Significant reduction of irido-corneal angels and excessive vault was observed, the lens was removed and exchanged for a shorter lens on (b)(6) 2018 and the problem was resolved.
 
Manufacturer Narrative
Additional information: device evaluation: the lens was returned in a micro centrifuge vial with moisture on the lens.Visual inspection found the haptic torn.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7906913
MDR Text Key121637062
Report Number2023826-2018-01438
Device Sequence Number1
Product Code QCB
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model NumberVTICMO13.2
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/18/2018
Initial Date FDA Received09/25/2018
Supplement Dates Manufacturer Received03/15/2019
Supplement Dates FDA Received03/16/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age22 YR
-
-