Model Number 8826545006
Device Problem Malposition of Device (2616)
Patient Problem No Code Available (3191)
Event Date 08/28/2018
Event Type  Injury  
Manufacturer Narrative
No product has been returned for evaluation as product remains in-situ.No product malfunction reporter.No radiographic images provided therefore alleged malfunction could not be confirmed.A review of the information provided identified that potential anatomical challenges may have caused or contributed to the malplacement.Labeling review: ".Use lateral fluoroscopy to properly manage the k-wire during pedicle preparation to ensure proper placement and avoid anterior advancement of the k-wire." ".The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient." ".As with any major surgical procedures, there are risks involved in orthopedic surgery.Infrequent operative and postoperative complications that may result in the need for additional surgeries include: early or late infection; damage to blood vessels, spinal cord or peripheral nerves; pulmonary emboli; loss of sensory and/or motor function; impotence; and permanent pain and/or deformity.Rarely, some complications may be fatal.Potential risks identified with the use of this system, which may require additional surgery, include: bending, fracture or loosening of implant component(s), loss of fixation, nonunion or delayed union, fracture of the vertebra, neurological, vascular or visceral injury.".
 
Event Description
On (b)(6) 2018 a patient underwent a posterior lumbar interbody fusion at l4/5 levels.As per reporter the surgeon noted difficulty determining proper position of screws.On (b)(6) 2018 the surgeon reported the l5 left screw breached into the vertebral canal.No alleged product malfunction.No post-operative adverse symptoms nor revision procedure is planned at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PRECEPT SPINAL SYSTEM
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
NUVASIVE INC.
7475 lusk blvd
san diego CA 92121
Manufacturer (Section G)
NUVASIVE
7475 lusk blvd
san diego CA 92121
Manufacturer Contact
george panfili
7475 lusk blvd
san diego, CA 92121
MDR Report Key7908131
MDR Text Key121706512
Report Number2031966-2018-00145
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00887517312358
UDI-Public887517312358
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K122352
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number8826545006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/29/2018
Initial Date FDA Received09/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1