While using a disposable pressure transducer (dpt), the diastolic blood pressure (dbp), systolic blood pressure (sbp) and mean arterial pressure (map) were observed to be lower than expected.It was indicated that there was no alarm or error message displayed.The user did not remember the displayed value.The issue was solved by using another transducer.The patient was not monitored with any other devices and no treatment was given.There was no allegation of patient injury.
|
The reported event of issue with pressure reading was not confirmed.The dpt zeroed and sensed pressure accurately on a pressure monitor.Pressure readings were stable during 8 hours of an output drift test.Electrical testing showed that both input and output impedances were within specifications.Zero-offset also met specification.No leakage or occlusion was detected from the dpt kit during a pressure test.In addition, no visible damage was observed from the dpt kit.The reported event could not be confirmed or replicated during the analysis, as the device responded appropriately during functional testing.There was no evidence of a manufacturing nonconformance.Poor dynamic response can be caused by air bubbles, clotting, and excessive lengths of tubing, excessively compliant pressure tubing, small bore tubing, loose connections, or leaks.The assembly may be tested for dynamic response by observing the pressure waveform on an oscilloscope or monitor.Bedside determination of the dynamic response of the catheter, monitor, kit and transducer system is done after the system is flushed, attached to the patient, zeroed and calibrated.A square-wave test may be performed by pulling the snap tab device and releasing quickly.Pressure readings can change quickly and dramatically because of loss of proper calibration, loose connection, or air in the system.Pressure readings should correlate with the patient¿s clinical manifestations.It is unknown if user or procedural factors contributed to the stated event.In this event, there was no patient compromise noted.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|