• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: POSEY PRODUCTS LLC POSEY BED 8060; PATIENT BED WITH CANOPY/RESTRAINTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

POSEY PRODUCTS LLC POSEY BED 8060; PATIENT BED WITH CANOPY/RESTRAINTS Back to Search Results
Model Number 8060
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Inspection found a zipper element missing, which can allow for the unintentional opening of the canopy window.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.Instructions for use(ifu) were reviewed and found to provide adequate instructions and warning for safe and effective use of the device.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
 
Event Description
Customer reported the canopies have zippers that are damaged.Customer provided limited information.The date the issue was discovered is unknown and no patient incident or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POSEY BED 8060
Type of Device
PATIENT BED WITH CANOPY/RESTRAINTS
Manufacturer (Section D)
POSEY PRODUCTS LLC
5635 peck road
arcadia CA 91006
Manufacturer Contact
william hincy
5635 peck road
arcadia, CA 91006
6264433143
MDR Report Key7917190
MDR Text Key122008749
Report Number2020362-2018-00135
Device Sequence Number1
Product Code OYS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113355
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 08/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number8060
Device Catalogue Number8060
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2018
Initial Date FDA Received09/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-