MAUDE Adverse Event Report: STRYKER GMBH RECONSTRUCTION PLATE, STRAIGHT HOLES 7 LENGTH 84MM FOR SCREWS Ø3.5/4.0MM; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH RECONSTRUCTION PLATE, STRAIGHT HOLES 7 LENGTH 84MM FOR SCREWS Ø3.5/4.0MM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 430107
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problems Fall (1848); Fracture, Arm (2351)
Event Date 08/01/2018
Event Type  Injury  
Manufacturer Narrative
Device will not be returned.If additional information becomes available it will be provided on a supplemental report.Explanted device are not available.
 
Event Description
The customer reported that the surgeon implanted a recon plate instead of a waisted compression plate into the wrist of a patient.The items are positioned in similar places on the small fragment set tray and have a similar appearance when viewed side by side and as such the wrong plate was handed to the surgeon in error.The customer reported that the patient had a fall after plating and presented with a bent arm.Patient required revision surgery.
 
Manufacturer Narrative
The reported event that reconstruction plate, straight holes 7 length 84mm for screws 3.5/4.0mm was alleged of 'wrong device used' could be confirmed only according to the given feedback from the hospital, despite that the product was not returned.Based on investigation, the root cause was attributed to be user related.The failure was caused by an user error as the wrong plate was taken out of the tray and used (a recon plate instead of a waisted compression plate).Original statement from our medical expert: more detailed information should be known in order to be able to give a sound answer to this occurrence.In general i would think that no major problems are associated with this wrong choice.In some cases the plates have to be bend for final exact fixation anyway but in the end the doctor and the other hcp involved are responsible which device they use.A review of the device history was not possible because the lot number was not communicated.Note that this only happened in this particular hospital, therefore the relevant hospital staff needs to undergo a training (¿user ¿ customer - learning gaps errors; lack of knowledge skills¿) in order to eliminate such problems.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
The customer reported that the surgeon implanted a recon plate instead of a waisted compression plate into the wrist of a patient.The items are positioned in similar places on the small fragment set tray and have a similar appearance when viewed side by side and as such the wrong plate was handed to the surgeon in error.The customer reported that the patient had a fall after plating and presented with a bent arm.Patient required revision surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECONSTRUCTION PLATE, STRAIGHT HOLES 7 LENGTH 84MM FOR SCREWS Ø3.5/4.0MM
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key7924063
MDR Text Key122301798
Report Number0008031020-2018-00699
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
PMA/PMN Number
K000636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue Number430107
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/12/2018
Initial Date FDA Received10/01/2018
Supplement Dates Manufacturer Received11/08/2018
Supplement Dates FDA Received12/06/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-