TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4415
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported that the heater cooler was not switching from the small tank to the large tank.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The manufacturer's technical support specialist verified the reported complaint.Valve five was found to be frozen so he replaced it.He also replaced valve four as a precaution.The unit operated to the manufacturer's specifications.The suspect parts were returned to the manufacturer for further evaluation.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) observed that the inside of the valve contained a large amount of corrosion preventing it from operating correctly.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7927008
MDR Text Key122458135
Report Number1828100-2018-00503
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/01/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/07/2018
Initial Date FDA Received10/02/2018
Supplement Dates Manufacturer Received10/29/2018
11/30/2018
Supplement Dates FDA Received11/20/2018
12/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-