• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 3116
Device Problems Entrapment of Device (1212); Grounding Malfunction (1271); Migration or Expulsion of Device (1395); Energy Output Problem (1431); Insufficient Information (3190)
Patient Problems Nausea (1970); Foreign Body In Patient (2687)
Event Date 07/02/2014
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 435135, serial# (b)(4), implanted: (b)(6) 2008, explanted: (b)(6) 2014, product type: lead.Product id: 435135, serial# (b)(4), implanted: (b)(6) 2008, explanted: (b)(6) 2014, product type: lead; other relevant device(s) are: product id: 435135, serial/lot #: (b)(4), ubd: 17-dec-2009, udi#: (b)(4); product id: 435135, serial/lot #: (b)(4), ubd: 13-dec-2009, (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer (con) regarding a patient who was implanted with a neurostimulator for gastric stimulation.It was reported that the patient needed a complete replacement since their leads had wrapped up in the scar tissue in their stomach.The patient had a pyloroplasty done and their nausea started getting worse.They had their device looked at and there was no signal getting to the leads, so it was decided to replace both the leads and the implanted neurostimulator (ins).No further complications were reported or anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins).It was reported her battery was getting and she needed it replaced.The patient was transferred to patient services.There were no further complications that have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7931213
MDR Text Key122542981
Report Number3004209178-2018-22169
Device Sequence Number1
Product Code LNQ
UDI-Device Identifier00643169174993
UDI-Public00643169174993
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2014
Device Model Number3116
Device Catalogue Number3116
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/02/2018
Initial Date FDA Received10/03/2018
Supplement Dates Manufacturer Received10/02/2018
Supplement Dates FDA Received10/04/2018
Date Device Manufactured03/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age47 YR
-
-