• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4416
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/20/2017
Event Type  malfunction  
Manufacturer Narrative
It was reported that while removing the ice, compresses were found on the bottom covering the water outlet.
 
Event Description
It was reported that the arterial circuit of the heater cooler was not circulating.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.Per the manufacturer's subsidiary, the compresses were removed from the water outlet resolving the issue.The compresses were discarded and will not be returned for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7935950
MDR Text Key122808212
Report Number1828100-2018-00517
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number4416
Device Catalogue Number4416
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2018
Initial Date FDA Received10/04/2018
Supplement Dates Manufacturer Received11/26/2018
Supplement Dates FDA Received11/28/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-