AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) GREENLIGHT MOXY FIBER OPTIC; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0010-2400
Device Problems Fracture (1260); Detachment of Device or Device Component (2907); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/18/2018
Event Type  malfunction  
Manufacturer Narrative
Product evaluation: failure analysis for fiber 0010-2400-739a-(b)(4): the glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge; the fiber proximal to fracture can rotate independently of metal cap; the distal part of glass cap and metal cap are detached and not returned; the glass cap exhibits severe devitrification at output window; the outer flow tubing open end exhibits minor scratch marks and signs of melting; the connector cone, segments, and tabs appear in good condition and secured; the fiber connector passed the signature verification fixture test.Based on device analysis, the potential for forward firing may exist.Probable root cause: based on the device analysis, the product complaint "fiber connector broke in the fiber port" could not be confirmed.The probable root cause of complaint is: not confirmed ¿ returned.Additional information: fiber cap detached observed.The probable root cause was determined based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.
 
Event Description
It was reported that during a benign prostatic hyperplasia (bph) procedure a defective greenlight moxy fiber was noted.The fiber was exchanged however it was noted that the fiber "hub" broke in the fiber port which is being interpreted as the fiber connector not functioning in the fiber port.The procedure was completed with bipolar resection (turp).No injury reported.This report is for the second fiber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7943715
MDR Text Key123370828
Report Number2937094-2018-01052
Device Sequence Number1
Product Code GEX
UDI-Device Identifier00878953005515
UDI-Public00878953005515
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/21/2019
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number739A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Initial Date Manufacturer Received 10/02/2018
Initial Date FDA Received10/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-