Model Number 6379
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Patient Involvement (2645)
Event Date 09/13/2018
Event Type  malfunction  
Manufacturer Narrative
Per field service representative (fsr) the flow would intermittently be displayed and then just read.00 flow.The unit has reached its end of service life and replacement parts are no longer available.The user facility has been advised to remove the unit from clinical use.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, the unit was not reading a flow.There was no patient involvement.
 
Manufacturer Narrative
The reported complaint has been confirmed.During laboratory analysis, the product surveillance technician (pst) observed there to be no flow readings due to a corroded cable connector on the internal flow meter.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS CENTRIFUGAL SYSTEM
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS - SARNS CENTRIFUGAL SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7943814
MDR Text Key123259499
Report Number1828100-2018-00520
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K950739
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number6379
Device Catalogue Number6379
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/13/2018
Initial Date FDA Received10/08/2018
Supplement Dates Manufacturer Received11/01/2018
Supplement Dates FDA Received11/06/2018
Is This a Reprocessed and Reused Single-Use Device? No