TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
|
Back to Search Results |
|
Model Number 816571 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the user facility's biomedical engineer, after consulting with the field service representative (fsr) it was determined that this is a common message displayed when performing the safety check with this model of roller pump.
|
|
Event Description
|
It was reported that during preventive maintenance (pm) of the device, the roller pump displayed a "contact your service provider" message instead of the expected "pump jam" error message.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Search Alerts/Recalls
|
|
|