TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II); CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING Back to Search Results
Model Number 4416
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/24/2017
Event Type  malfunction  
Event Description
It was reported that the arterial does not circulate on the heater cooler.No other details regarding the nature of this event were provided.
 
Event Description
Further information was received that the unit would not recirculate.
 
Manufacturer Narrative
The reported complaint was confirmed.The technician from the manufacturer's subsidiary replaced the threaded polyvinyl chloride (pvc) and the barbed tee reducer to resolve the issue.No parts will be returned to the manufacturer for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS TEMP. CONTROL AND MONITOR SYSTEM (TCM II)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-SARNS TCM II HEATING AND COOLING
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7944354
MDR Text Key123259355
Report Number1828100-2018-00523
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number4416
Device Catalogue Number4416
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2018
Initial Date FDA Received10/08/2018
Supplement Dates Manufacturer Received11/26/2018
Supplement Dates FDA Received11/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-