Catalog Number 05122279001 |
Device Problem
Low Test Results (2458)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/19/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).
|
|
Event Description
|
The customer complained of discrepant results for 1 patient tested for hemoglobin and hct on a cobas b 123 instrument compared to the beckman coulter hematology analyzer.The patient was initially tested on the b 123 instrument with a hemoglobin result of 4.1 g/dl and an hct of 16%.These results were reported outside of the laboratory where the physician treated the patient with a transfusion of 2 units of erythrocyte concentrates.The patient was tested by the beckman coulter analyzer and the hemoglobin result was 8.2 g/dl and the hct was 24.9% so the transfusion was not necessary.There was no allegation that an adverse event occurred due to the transfusion.The sensor cartridge lot number and expiration date were not provided.Qc values for hemoglobin were acceptable.No alarms were produced by the b 123 instrument.
|
|
Manufacturer Narrative
|
The customer indicated that the transfusion was performed on the same day as the erroneous results were obtained (b)(6) 2018) in the afternoon (exact time not indicated).The provided printout of comparison results from cobas b 123 sn (b)(4)is dated from (b)(6)2018 at 00:48; and the printout of measurement performed on the beckman coulter is dated from (b)(6) 2018 at 01:43.Further clarification regarding the dates of the measurements were requested, but not provided.Hemoglobin and hct are measured independently from each other on the b 123 system.Both results are decreased in comparison to the coulter results.The results obtained are consistent with a preanalytical sample handling problem, such as sedimentation in a poorly mixed whole blood sample.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Search Alerts/Recalls
|