Catalog Number ATS45
Device Problems Failure to Advance (2524); Failure to Fire (2610)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/01/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r58n5z.Investigation summary
=
> the analysis results found that the ats45 device was received with the firing mechanism damaged and with a tr45w cartridge loaded in the device.The reload was received partially fired 1/10 and with the reload lock out spring damaged.In addition cartridge b, 6r45b, batch r58u4r, was received, fully fired, with the lockout normal.No functional test could be performed due to the condition of the device.The device was disassembled to verify the condition of the internal components and the firing trigger teeth were found broken.The damage to the reload lockout spring is consistent with damage observed when the firing cycle is started, interrupted, released, and restarted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.Please reference the instruction for use for more information.It should be noted that a 100% inspections takes place during manufacturing to ensure the device meets the require specifications; in addition, a sample of the batch is inspected at fgqa.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during a laparoscopic appendectomy procedure that the staff loaded the device the surgeon attempted to fire but the device locked out.A few of the staples deployed but the knife blade did not advance through any tissue.The device was removed, and a second device was opened.Another like device was used to complete the procedure.There were no adverse consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS FLEX ARTICNG LNR CUTR 45MM
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7951620
MDR Text Key123374085
Report Number3005075853-2018-13415
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036001072
UDI-Public10705036001072
Combination Product (y/n)N
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue NumberATS45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2018
Initial Date Manufacturer Received 09/12/2018
Initial Date FDA Received10/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-