• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problem Circuit Failure (1089)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/22/2018
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device had a broken crystal.Off-site hybrid analysis confirmed the field reported no telemetry failure.The output of the crystal oscillator was not able to produce the proper 32khz signal causing a no telemetry condition.X-ray inspection of the crystal oscillator component revealed a broken quartz crystal element.The device was fully functional with nominal system current drain, reference voltages, and regulated supply voltages after the fault crystal oscillator component was replaced with a known-good one.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the implantable cardiac monitor (icm) was explanted due to system upgrade.Analysis of the device was performed and the device tested out of specification.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7951678
MDR Text Key123234402
Report Number9614453-2018-03489
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169732216
UDI-Public00643169732216
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/14/2018
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2018
Initial Date Manufacturer Received 06/06/2018
Initial Date FDA Received10/10/2018
Date Device Manufactured02/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-