• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number 990063-020
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hemoptysis (1887); Great Vessel Perforation (2152); Blood Loss (2597)
Event Date 10/08/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, it was noted that the physician had a difficult time finding the right inferior pulmonary vein (ripv) and the mapping catheter may have entered into an intermediate small right pulmonary vein.Soon after, the patient started to cough and the nurse noticed blood was being expectorated.The physician stopped the catheter manipulation to check on the patient and consulted with vascular experts.Medication was administered to reverse the bleeding.The bleeding stopped quickly and the patient stopped coughing.The case was aborted.The patient was ok and a scan was performed which revealed some bleeding around the small intermediate right pulmonary vein that was stable.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed at least 5 injections were performed with balloon catheter 2af283 with lot number 75268 with system notice 50032 ¿the safety system detected a compromised outer vacuum triggered¿ on one injection on the date of event.This system notice is not related to the reported clinical issue.Clinical issues were encountered during the procedure.The mapping catheter was not returned for investigation.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7951746
MDR Text Key123237138
Report Number9612164-2018-02695
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/30/2020
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number215901479
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/08/2018
Initial Date FDA Received10/10/2018
Supplement Dates Manufacturer Received10/11/2018
Supplement Dates FDA Received11/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Weight76
-
-