• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ETS FLEX ARTICNG LNR CUTR 45MM; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number ATS45
Device Problems Failure to Advance (2524); Failure to Fire (2610)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r58n5z.Investigation summary
=
> the analysis results found that the ats45 device a was received with no apparent damaged and with a 6r45b cartridge loaded on the device.The returned reload was partially fired 1/10.The returned device was tested for functionality with a test reload and it fired, cut, and all the staples formed as intended.The device fired without any difficulties, the staple line was complete, and the staples were noted to have the proper b-formed shape.The condition of the reload indicates that the device's firing cycle was interrupted.When firing the device make sure that the firing stroke is completed.Do not partially fire the device.Fire the device by squeezing the firing trigger completely until it rests on the closing trigger.Once the firing cycle has been initiated, it must be completed.If re-initiation of firing is resumed, the device will lockout.Firing through the lockout mechanism will break the device.The batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during a laparoscopic appendectomy procedure that the staff loaded the device the surgeon attempted to fire but the device locked out.A few of the staples deployed but the knife blade did not advance through any tissue.The device was removed, and a second device was opened with the same outcome.Another device was used to complete the procedure.There were no adverse consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETS FLEX ARTICNG LNR CUTR 45MM
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7952158
MDR Text Key123371960
Report Number3005075853-2018-13426
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10705036001072
UDI-Public10705036001072
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2023
Device Catalogue NumberATS45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/11/2018
Initial Date Manufacturer Received 09/12/2018
Initial Date FDA Received10/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-