Model Number 801764 |
Device Problem
Power Problem (3010)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's subsidiary, there was a bad electrical connection, false contact in the network input circuit.
|
|
Event Description
|
It was reported that the heart lung machine (hlm) would only operate in battery mode.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.The repair technician from the manufacturer's subsidiary found that the electrical connection had false contact in the network input circuit.He repaired that connection.There will be no parts returned to the manufacturer for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|