• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801764
Device Problem Power Problem (3010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2017
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's subsidiary, there was a bad electrical connection, false contact in the network input circuit.
 
Event Description
It was reported that the heart lung machine (hlm) would only operate in battery mode.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.The repair technician from the manufacturer's subsidiary found that the electrical connection had false contact in the network input circuit.He repaired that connection.There will be no parts returned to the manufacturer for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7952399
MDR Text Key123259593
Report Number1828100-2018-00545
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801764
Device Catalogue Number801764
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2018
Initial Date FDA Received10/10/2018
Supplement Dates Manufacturer Received11/29/2018
Supplement Dates FDA Received11/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-