Model Number 0010-2400
Device Problems Fracture (1260); Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2016
Event Type  malfunction  
Manufacturer Narrative
Analysis of the returned device revealed that the glass cap shows a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge.The fiber proximal to fracture can rotate independently of outer flow tubing.The glass cap evel edge and the metal cap are not aligned, the glass cap is protruding from the metal cap.The glass cap exhibits severe devitrification at the output window.The metal cap severe charred detritus on surface.The outer flow tubing open end exhibits minor scratch marks and minor contamination most like caused by biological material.The white tubing was noted to be melted but the fiber optic is intact.Based on device analysis, the potential for forward firing may exist.The probable cause of the failure is operational context.The cause was determined based on heat accumulation.Cap wear was accelerated due to anatomical/procedural factors (tissue contact and technique) encountered during the procedure which would limit the performance of the fiber.The identified issues noted above may activate the fiberlife function which would modulate the power showing a pulsing beam or system would be placed into standby mode.
 
Event Description
Analysis of the return device revealed a circumferential fracture on the distal side of fiber/cap fusion zone at the bevel edge.Based on the device analysis, the potential for forward firing may exist.There were no patient complications reported due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENLIGHT MOXY FIBER OPTIC
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
3070 orchard drive
san jose CA 95134
Manufacturer Contact
alyson harris
3070 orchard drive
san jose, CA 95134
4084563300
MDR Report Key7954754
MDR Text Key123369855
Report Number2937094-2018-01063
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 11/25/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2018
Device Model Number0010-2400
Device Catalogue Number0010-2400
Device Lot Number630A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Initial Date Manufacturer Received 10/09/2018
Initial Date FDA Received10/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/25/2016
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-