MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC HD 1000I SHEARS 36CM SHAFT; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HARHD36
Device Problems Break (1069); Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r92j5u.Investigation summary: the device was returned with the tissue pad damaged, melted and 100% present.The device was connected to a gen11 and the device did activate during functional testing.The device was disassembled to inspect the internal components and no anomalies were found, related to the reported event.In addition, when the device was disassembled and it was found that there was a cracked internal component; the y-link.Probable causes of tissue pad damage are applying pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back-cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
 
Event Description
It was reported that during a laparoscopic sigmoidectomy procedure, the device could not cut the tissue easily and the tissue pad was damaged.The target tissue was stiff and severely adhered.The device was used on the colon and rectum.The blade tip was not broken off and no pieces fell into the patient.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC HD 1000I SHEARS 36CM SHAFT
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key7960215
MDR Text Key123684549
Report Number3005075853-2018-13539
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036015055
UDI-Public10705036015055
Combination Product (y/n)N
PMA/PMN Number
K160752
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Catalogue NumberHARHD36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Initial Date Manufacturer Received 09/27/2018
Initial Date FDA Received10/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/28/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-