The customer complained of discrepant sodium (na) results for 1 patient tested on a cobas b 123 system compared to a beckman coulter analyzer.The na result from the beckman coulter analyzer at 1:59 p.M.Was 162 mmol/l.A sample was obtained at 1:59 p.M.For the cobas b 123 instrument and the na result at 2:09 p.M.Was 152 mmol/l.The customer performed comparison testing with the patient sample for troubleshooting purposes on 4 other cobas b123 instruments: cobas b 123 #1 at 2:11 p.M.Was 156 mmol/l, cobas b 123 #2 at 2:17 p.M.Was 159 mmol/l, cobas b 123 #3 at 2:31 p.M.Was 161 mmol/l, cobas b 123 #4 at 2:37 p.M.Was 161 mmol/l.The customer is concerned with the lower na result of 152 mmol/l from the b 123 instrument in question.There was no allegation that an adverse event occurred.Qc and calibration were acceptable.The na electrode lot number and expiration date were not provided.
|